- Large easy to read display of measurement value and temperature
- Automatic calibration
- Automatic or manual temperature compensation
- Energy saving switch (Auto-Off)
- Freeze the measured value on the display (HOLD)
- Auto-monitoring after switching on
- Battery power supply
- Available in complete suitcase sets with conductivity cell and calibration equipment
Characteristic
Available Models
A simple conductivity meter to meet high requirements, with only 5 buttons, parallel temperature measurement, automatic temperature compensation; 4-electrode measurement technique.
Rugged and waterproof pocket conductivity meter, battery-operated, with parallel temperature display, with integrated memory, equipped with GLP-required functions, automatic and manual temperature compensation with linear function, as well as non-linear function for natural and ultrapure waters according to DIN 38 404 or EN 27 888. Temperature compensation can be turned off. 20 °C or 25 °C can be selected as the reference temperature. The Cond 340i model additionally has an analog output and an RS 232 digital output, as well as an optional AC power supply.
Comparison of available models:
Technical data | Cond 315i | Cond 330i | Cond 340i |
Parameters | X, Ω, Sal, °C | X, TDS, Ω, Sal, °C | |
conductivity range resolution accuracy calibration cell constant reference temperature temperature compensation temperature coefficient |
0 to 500 mS/cm from 0,01 uS/cm 0,5% of measured value 1 pkt. 0,1; 0,475 20 lub 25 nLF(-5 to +105 °C) without compensation - |
0 to 500 mS/cm from 0,01 uS/cm for K=0,01 cm-1 0,5% of measured value 1 pkt. 0,01; 0,1; 0,475; 0,25-2,5 20 lub 25 Lin(-5 to +105 °C) NFL (-5 to +105 °C) without compensation 0,01 to 2,99 |
|
TDS range [ppt]. resolution accuracy calibration factor |
0 to 1999 1 +/- 1 mg/l 1 point 0,40 to 1,0 |
- - - - - |
|
resistance range resolution |
0,000 to 1999 MΩcm to 0,01 |
0,000 to 1999 MΩcm to 0,01 |
|
salinity range resolution |
0,0-70 0,1 |
/ | |
temperature range resolution accuracy |
-5,0 to +105,0 °C 0,1 +/-0,1 |
/ | |
memory GLP function inputs outputs power supply auto-off protection |
- - 8-pin connector - battery after 1 h from the last key press IP 67 |
800 yes 8-pin connector - battery after 1 h from the last key press IP 67 |
800 yes 8-pin connector RS 232 battery after 1 h from the last key press IP 67 |
Model | Parameter | Description |
---|---|---|
Cond 315i | X, Ω, Sal, oC | 8-pin plug, battery powered, IP 67 protection |
Cond 330i | X, TDS, Ω, SAL, oC | Memory 800, GLP, 8-pin plug, battery power, IP67 protection |
Cond 340i | X, TDS, Ω, SAL, oC | Memory 800, GLP, 8-pin plug, mains battery power, IP67 protection |