Stationary EC/TDS conductivity meter
Choose Models for offer request
Intended
- Temperature compensation: automatic or manual, temperature compensation can be turned off if the actual conductivity value is required
- Function to automatically determine the appropriate range for EC and TDS values
- LCD display
- Auto shut-off function to extend battery life
- Internal clock and date for tracking various time-dependent functions
- Built-in rechargeable battery with a life of up to 8 hours
Characteristic
Available Models
Technical specification:
Model | MW170 MAX | |
Range | EC | 0,00 to 29,99 uS/cm; 30,0 to 299,9 uS/cm; 300 to 2999 uS/cm; 3,00 to 29,99 mS/cm; 30,0 to 200,0 mS/cm; to 500,0 mS/cm, absolute conductivity |
TDS | 0,00 to 14,99 ppm (mg/L); 15,0 to 149,9 ppm (mg/L); 150 to 1499 ppm (mg/L); 1,50 to 14,99 g/L; 15,0 to 100,0 g/L (ppt); to 400,0 g/L Absolute TDS* (at a factor of 0.80) | |
Salinity | 0,0 to 400,0 %NaCl; 2,00 to 42,00 PSU; 0,00 to 80,00 g/L | |
°C | -20,0 to 120,0°C | |
Resolution | EC | 0,01 uS/cm / 0,1 uS/cm / 1 uS/cm / 0,01 mS/cm / 0,1 mS/cm |
TDS | 0,01 ppm / 0,1 ppm / 1 ppm / 0,01 g/L / 0,1 g/L | |
Salinity | 0,1 % NaCl / 0,01 PSU / 0,01 g/L | |
°C | 0,1°C | |
Accuracy | EC | ± 1% of reading / ± 0.05 uS/cm or 1 digit, whichever is greater |
TDS | ±1% of reading / ±0.03 ppm or 1 digit, whichever is greater | |
Salinity | ± 1% of reading | |
°C | ±0,5°C | |
Temperature compensation | ATC- automatic, from -5 to 100°C; MTC - manual, from -5 to 100°C; NO TC - no compensation | |
Calibration EC/TDS | Single cell factor calibration, 6 standards: 84 uS/cm, 1413 uS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm / single point offset: 0.00 uS/cm | |
Calibration of salinity | 1-point with solution for salinity calibration MA9066 | |
Sensor included | MA814DB/1 | |
Battery life | 8 h | |
Dimensions | 230 x 160 x 95 mm | |
Weight | 0,9 kg |
Model | Battery life |
---|---|
MW170 MAX | 8 h |
Check also