Demineralizers Technical Series
Choose Models for offer request
Intended
- Automatic and maintenance-free operation of the device
- Color LCD display
- Third class of water purity
- Possibility of connection to a dishwasher, autoclave, etc.
Characteristic
Available Models
TECHNICAL series deionizers are commonly used water treatment systems powered by tap water, which retain 96-99% of organic and inorganic pollutants dissolved in water. They are mainly used as a source of water for dishwashers, autoclaves, batteries, climatic and salt chambers, and for general laboratory activities.
Technical data:
- Power supply: tap water
- Possibility of connection to a dishwasher, autoclave, washer, climatic chamber, purification system, etc.
- Possibility of installing a UV sterilizer
- Third class of purity according to PN-EN ISO 3696:1999
Basic model automation:
- Color LCD display
- Conductometer measuring the conductivity and temperature of purified water
- Alarms: informing about the replacement of the mechanical and carbon filter
- Preview of service dates
- Built-in RS 232 connector for communication with a computer providing the possibility of individual adjustment of service frequencies and alarm levels
- Water supply pressure gauge
Automation of the model with C control:
- Color touch graphic display with Touch Panel function
- Conductometer measuring the conductivity and temperature of supply water, after reverse osmosis and ultra-pure in units of µS/cm or MOhm interchangeably
- Clock displaying the date and time
- Alarms: informing about the replacement of the mechanical and carbon filter
- Built-in RS 232 connector for communication with a computer providing the possibility of individual adjustment of service frequencies and alarm levels
- Water supply pressure gauge
- Disruption of pump operation at: low supply water pressure (lack of supply water) and full tank
Serwis:
Model |
Sediment prefilter | Module A2 | Radiator UV 254NM | Carbon filter GAC 10" |
Technical 5 |
+ | + | opcja | - |
Technical 5/C | + | + | opcja | - |
Technical 10 | + | + | opcja | - |
Technical 10/C | + | + | opcja | - |
Technical 20 | + | + | opcja | - |
Technical 20/C | + | + | opcja | - |
Technical 30 | + | + | opcja | - |
Technical 30/C | + | + | opcja | - |
Technical 40 |
+ | - | opcja | + |
Technical 40/C | + | - | opcja | + |
Technical 60 | + | - | opcja | + |
Technical 60/C | + | - | opcja | + |
Work time | 6 months* | 6 months* | 8500 h | 6 months* |
Cat no | EO-005-10 | EO-MA-12 | EUV-254-HLP | EW-001-10 |
* the life of the cartridge may change depending on the flow rate, its characteristics and the level and type of contamination of tap water.
Model | System performance [l/h] |
---|---|
Technical 5 | 5-7 |
Technical 5/C | 5-7 |
Technical 10 | 10-12 |
Technical 10/C | 10-12 |
Technical 20 | 20-22 |
Technical 20/C | 20-22 |
Technical 30 | 30-32 |
Technical 30/C | 30-32 |
Technical 40 | 40-42 |
Technical 40/C | 40-42 |
Technical 60 | 60-62 |
Technical 60/C | 60-62 |