Intended

  • Color LCD display
  • Third class of water purity
  • Possibility of connection to a dishwasher, autoclave, etc.

The TECHNICAL PLUS series deionizers have been created to work in laboratories that need water in the range of 0.06 µS/cm to 15-20 µS/cm. By adjusting the conductivity, the user can decide on the quality of water they want to obtain. This process allows for the optimization of operating costs. TECHNICAL PLUS systems are a source of water for autoclaves, dishwashers, analyzers, climatic cabinets, for the preparation of reagents, buffers, pH solutions, as well as for chemical analyses and syntheses.

Technical data:

• Power supply: tap water

• Possibility of connection to a dishwasher, autoclave, washer, climatic chamber, purification system, etc.

• Possibility of installing a UV sterilizer

• Third class of purity according to PN-EN ISO 3696:1999

Basic model automation:

• Color LCD display

• Conductometer measuring the conductivity and temperature of purified water

• Preview of service dates

• Built-in RS 232 connector for communication with a computer providing the possibility of individual adjustment of service frequencies and alarm levels

• Water supply pressure gauge

 

Automation of the model with C control:

• Color touch graphic display with Touch Panel function

• Conductometer measuring the conductivity and temperature of supply water, after reverse osmosis and ultra-pure in units of µS/cm or MOhm interchangeably

• Clock displaying the date and time

• Built-in RS 232 connector for communication with a computer providing the possibility of individual adjustment of service frequencies and alarm levels

• Water supply pressure gauge

• Disruption of pump operation at: low supply water pressure (lack of supply water) and full tank


Serwis:

Model
Sediment prefilter 5μm Module A2 Radiator UV 254nm Module 2xH7 Module H6
Technical 5 Plus  + + opcja + -
Technical 5/C Plus  + + opcja + -
Technical 10 Plus + + opcja - +
Technical 10/C Plus + + opcja - +
Technical 20 Plus
+ + opcja - +
Technical 20/C Plus + + opcja - +
Technical 30 Plus
+ + opcja - +
Technical 30/C Plus + + opcja - +
Technical 40 Plus
+ + opcja - +
Technical 40/C Plus + + opcja - +
Technical 60 Plus
+ + opcja - +
Technical 60/C Plus + + opcja - +
Work time 6 months* 6 months* 8500 h 2000 dm³** 5000 dm³**
Cat no EO-005-10 EO-MA-12 EUV-254-HLP EJ-2000-0 EJ-5000-0

* the life of the cartridge may change depending on the flow rate, its characteristics and the level and type of contamination of tap water.
** the volume of treated water depends on the quality of the feed water, the maximum amount of dissolved salts in the feed water - 1200 mg/l
Model System performance [l/h]
Technical 5 Plus 5-7
Technical 5/C Plus 5-7
Technical 10 Plus 10-12
Technical 10/C Plus 10-12
Technical 20 Plus 20-22
Technical 20/C Plus 20-22
Technical 30 Plus 30-32
Technical 30/C Plus 30-32
Technical 40 Plus 40-42
Technical 40/C Plus 40-42
Technical 60 Plus 60-62
Technical 60/C Plus 60-62