EN PL

Turbidity meter TB 300 IR

Choose Models for offer request

Intended

  • iso-9001-copy
  • shape-2601
  • shape-2680
  • bad_wlasc
  • iso-9001-copy
  • shape-2601
  • shape-2680
  • bad_wlasc
  • EN ISO 7027 compliant: 90° infrared light nephelometric method standard
  • Infrared light source: Clear and colored samples can be measured
  • 01-1100 NTU range with auto-range: automatic detection of the measuring range – no manual switching required
  • ±2% readout accuracy between 0.01 and 500 NTU: high resolution in low turbidity range
  • 1000 data sets + RS 232 interface: data recording and data transmission
  • Acid- and solvent-resistant membrane keyboard: Laboratory operation
  • Standard T-Cal calibration kit: automatic adjustment of the measuring range
  • Multi-language graphic display: user-friendly operation
  • Free software updates available from the manufacturer 's website www.lovibond.com

The TB 300 IR turbidimeter from Lovibond is a laboratory device for measuring the turbidity (turbidimetry) of liquids. Turbidity is measured according to EN ISO 7027 using the nephelometric method – measuring diffused light at a 90° angle. The infrared light source allows you to measure both colorless and colored samples. Autorange allows direct turbidity measurement between 0.01 and 1100 NTU.

The device is equipped with a large graphic display with a choice of several languages, internal memory for 1000 data sets, a real-time clock and an RS 232 interface. The keyboard is membrane, resistant to acids and solvents. Software updates (including language versions) are available free of charge from the www.lovibond.com manufacturer's website .

Technical specifications

Parameter

Value

Manufacturer

Lovibond

Model / Part No.

TB 300 IR / 194000-B

Meter Type

Stationary

Measurement principle

Nephelometric (diffused light at a 90° angle)

Standard

EN ISO 7027

Light source

Infrared (IR)

Measuring range

0.01-1100 NTU (autorange)

Accuracy

±2% reading or ±0.01 NTU (greater) between 0.01 and 500 NTU; ±5% readout in the range of 500-1000 NTU

Matching litter boxes

Round, 24 mm

Calibration

Standard T-Cal kit

Display

Graphic, multilingual

Keyboard

Membrane-resistant, acid- and solvent-resistant

Internal Memory

1000 datasets

Interface

RS 232

Clock

Real-time

Auto Shutdown

Yes

Power supply

7 nickel-cadmium batteries

Working conditions

5-40°C, 30-90% RH (non-condensing)

Dimensions

265 × 195 × 70 mm

Weight

1000 g (with batteries)

Application of TB 300 IR Turbidity Meter

The TB 300 IR turbidity meter is designed to measure the turbidity of liquids using the nephelometric method. Compliance with EN ISO 7027 defines that the measurement is performed with diffused light at an angle of 90° with an infrared light source.

The infrared light source allows you to measure both colourless and coloured samples, which is an important feature that distinguishes this type of device from turbidity meters that use white light.

Autorange eliminates the need for manual range switching – the device selects the optimal range for the sample to be measured over the full range of 0.01 to 1100 NTU. The accuracy is ±2% of the reading (or ±0.01 NTU, whichever is greater) in the range of 0.01-500 NTU, and ±5% of the reading in the range of 500-1000 NTU.

The standard T-Cal calibration kit provides automatic adjustment of the overall measuring range. The internal memory for 1000 data sets allows the results to be stored directly in the device, and the RS 232 interface allows data transfer. The acid- and solvent-resistant membrane keyboard makes it easy to operate. A large graphic display with a choice of several languages and a user-friendly user manual make the device easy to use.

Advantages of the TB 300 IR turbidity meter

  • EN ISO 7027 compliant: 90° infrared light nephelometric method standard
  • Infrared light source: Clear and colored samples can be measured
  • 01-1100 NTU range with auto-range: automatic detection of the measuring range – no manual switching required
  • ±2% readout accuracy between 0.01 and 500 NTU: high resolution in low turbidity range
  • 1000 data sets + RS 232 interface: data recording and data transmission
  • Acid- and solvent-resistant membrane keyboard: Laboratory operation
  • Standard T-Cal calibration kit: automatic adjustment of the measuring range
  • Multi-language graphic display: user-friendly operation
  • Free software updates available from the manufacturer 's website www.lovibond.com

Check the availability of the TB 300 IR Lovibond turbidity meter and request a quote via the request for quotation form.

Model Battery
TB 300 IR yes